Mon Feb 25 15:30:29 2013, yuta, Update, Alignment, Y arm locked, both colors
|
Mon Feb 25 17:58:28 2013, rana, Update, Alignment, Y arm locked, both colors
|
Mon Feb 25 22:42:32 2013, yuta, Summary, Alignment, current IFO situation
|
Mon Feb 25 21:25:14 2013, yuta, Update, Alignment, both arms locked in IR
|
Tue Feb 26 01:49:27 2013, Jenne, Update, Locking, PRMI locked
|
Tue Feb 26 02:04:51 2013, Koji, Update, Locking, PRMI locked
|
Tue Feb 26 11:55:11 2013, Manasa, Update, Locking, FPMI locked
|
Tue Feb 26 15:32:47 2013, Jenne, Update, Locking, FPMI locked
|
Wed Feb 27 01:09:53 2013, yuta, Update, Locking, PRMI sideband/carrier locked 6x
|
Wed Feb 27 02:44:39 2013, Jenne, Update, Locking, DRFPMI flashes
|
Wed Feb 27 20:10:43 2013, rana, Update, Locking, DRFPMI flashes
|
|
Message ID: 8170
Entry time: Tue Feb 26 11:55:11 2013
In reply to: 8166
Reply to this: 8171
|
Author: |
Manasa |
Type: |
Update |
Category: |
Locking |
Subject: |
FPMI locked |
|
|
[Yuta, Manasa]
FPMI locked.
Alignment and lock acquisition
1. Use ITMY/ETMY to maximize green transmission in Y arm (to get closer to arm alignment).
2. Use TT1/TT2 to align arms to IR to the Y-arm (Maximum TRY = 0.84)
3. Use BS/ITMX/ETMX to align IR to the X-arm.
4. Use POY11_I and POX11_I to lock Yarm and Xarm.
5. Use ASDC to lock MI.
Modified
1. X-arm trans camera settings changed.
2. Blocked stray green from reaching TRX PD and camera.
P.S. TRX seems to be moving less on the camera because of the oplevs centered and working from last night.
Plan
Lock green to X arm.
|