40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Feb 25 15:30:29 2013, yuta, Update, Alignment, Y arm locked, both colors 
    Reply  Mon Feb 25 17:58:28 2013, rana, Update, Alignment, Y arm locked, both colors 
       Reply  Mon Feb 25 22:42:32 2013, yuta, Summary, Alignment, current IFO situation IPANGdrift.png
    Reply  Mon Feb 25 21:25:14 2013, yuta, Update, Alignment, both arms locked in IR QUAD1_1045890588.png
       Reply  Tue Feb 26 01:49:27 2013, Jenne, Update, Locking, PRMI locked 
          Reply  Tue Feb 26 02:04:51 2013, Koji, Update, Locking, PRMI locked 
             Reply  Tue Feb 26 11:55:11 2013, Manasa, Update, Locking, FPMI locked 
                Reply  Tue Feb 26 15:32:47 2013, Jenne, Update, Locking, FPMI locked 
             Reply  Wed Feb 27 01:09:53 2013, yuta, Update, Locking, PRMI sideband/carrier locked 6x
                Reply  Wed Feb 27 02:44:39 2013, Jenne, Update, Locking, DRFPMI flashes 
                   Reply  Wed Feb 27 20:10:43 2013, rana, Update, Locking, DRFPMI flashes 
Message ID: 8157     Entry time: Mon Feb 25 15:30:29 2013     Reply to this: 8158   8162
Author: yuta 
Type: Update 
Category: Alignment 
Subject: Y arm locked, both colors 

[Koji, Yuta]

We aligned Y arm to Y green and tweaked TT1/TT2 to get IR locked in Y arm.

Alignment procedure:
  1. Align ETMY/ITMY to maximize TEM00 green transmission to PSL table. We reached ~240 uW.

  2. Aligned PRM and TT2 so that PRM reflected beam go through FI and get ITMY-PRM cavity flashing. This is to coarsely align input pointing to Y arm. After this alignment, we got tiny Y arm flash. Input pointing to IPANG QPD was lost.

  3. Aligned TT1/TT2 to maximize TRY in TEM00. We reached ~0.92.

Failed procedure:

  I was struggling with finding Y arm flash. I was using IPPOS/IPANG as input pointing reference, and slider values (C1:SUS-(ITMY|ETMY)_(PIT|YAW)_COMM) or OSEM values (C1:SUS-(ITMY|ETMY)_SUS(PIT|YAW)_IN1) before pumping for Y arm alignment reference. But it was a lot more easier if Y arm is aligned to green and having Yarm cavity axis assured.

Next:
  - X arm flash in IR
  - Steer X end green
  - If X arm or AS looks bad, adjust IR input pointing and Y arm alignment. We have to steer Y end green afterwards.