I undertook the investigation of the AS55 PD. I found the PD is not broken.
I tested the PD on the PD test bench and it works just fine.
I attatched the characterization result as there has been no detailed investigation of this PD as far as I remember.
The transimpedance gain at 55MHz is 420Ohm, and the shotnoise intercept current is 4.3mA. |