Message ID: 8092     Entry time: Fri Feb 15 21:22:29 2013     In reply to: 8091
Author: yuta 
Type: Update 
Category: Electronics 
Subject: AS55 replaced with POP55 PD 

I temporarily replaced AS55 PD with PD labeled "POP55(POY55)".
I think POP55 is working because I could lock MI with this PD using AS55_Q_ERR as an error signal. I rotated I/Q phase (C1:LSC-AS55_PHASE_R) to 70 deg by minimizing ASDC during MI lock.

POP55 PD was freely sitting on the ITMX table.
I will leave AS55 PD at free space of the AP table. Someone, please look into it.