Message ID: 8065     Entry time: Tue Feb 12 00:14:00 2013     Reply to this: 8093
Author: yuta 
Type: Update 
Category: Locking 
Subject: PRM coil balancing 

We tried to lock half-PRC tonight, but we couldn't. Why?? I could lock yesterday.
It locks for ~ 1 sec, but it beam spot motion freaks out mainly in yaw.
I tried to balance PRM coils, but oplev beam was clipped by MMT1......

What I did:
  1. Found elog #5392 and found F2P_LOCKIN.py

  2. Modified F2P_LOCKIN.py because LOCKIN channel names are some how changed like this;

LOCKIN1_I -> LOCKIN1_DEMOD_I
LOCKIN1_Q -> LOCKIN1_DEMOD_Q
LOCKIN1_SIG -> LOCKIN1_DEMOD_SIG

  3. Running

/opt/rtcds/caltech/c1/scripts/SUS/F2P_LOCKIN.py -o PRM

  should adjust (UL|UR|LR|LL)COIL_GAINs by putting some gain imbalance and shaking the mirror in different frequencies. It uses LOCKIN to OL(PIT|YAW).

  4. Since there was no PRM oplev beam coming out from the vacuum, I quickly looked into BS-PRM chamber. Oplev beam was clipped by MMT1. If I adjust PRM slider values to avoid clipping, the beam will be clipped by mirrors on oplev table. What happened to the PRM oplev?

  5. I also made bunch of /opt/rtcds/userapps/trunk/sus/c1/medm/templates/SUS_SINGLE_LOCKIN(1|2)_DEMOD_(I|Q|SIG).adl because there were missing screens.

Next:
 We need to restore the PRM oplev and balance the coils. See, also, elog #7679