Wed Feb 6 20:19:52 2013, Manasa, Update, Optics, G&H and LaserOptik mirrors
|
Thu Feb 7 14:10:25 2013, Manasa, Update, Optics, LaserOptik - AR Reflectivity - Bad data 
|
Fri Feb 8 21:14:52 2013, Manasa, Update, Optics, G&H - AR Reflectivity 
|
Fri Feb 8 22:49:31 2013, Koji, Update, Optics, G&H - AR Reflectivity
|
Fri Feb 8 23:04:40 2013, Manasa, Update, Optics, G&H - AR Reflectivity
|
Sat Feb 9 19:34:34 2013, rana, Update, Optics, G&H - AR Reflectivity
|
Mon Feb 11 19:55:47 2013, Manasa, Update, Optics, G&H - AR Reflectivity
|
Wed Feb 13 09:28:56 2013, Steve, Update, Optics, G&H - HR plots
|
|
Message ID: 8051
Entry time: Sat Feb 9 19:34:34 2013
In reply to: 8045
|
Author: |
rana |
Type: |
Update |
Category: |
Optics |
Subject: |
G&H - AR Reflectivity |
|
|
Use the trick I suggested:
Focus the beam so that the beam size at the detector is smaller than the beam separation. Use math to calculate the beam size and choose the lens size and position. You should be able to achieve a waist size of < 0.1 mm for the reflected beam. |