Message ID: 8025     Entry time: Thu Feb 7 17:10:11 2013     In reply to: 8022     Reply to this: 8029
Author: Koji 
Type: Summary 
Category: General 
Subject: PRC/arm mode matching calculations 

Quote:

I left out the current situation (PR2/3 with -600 RoC) and the case where only PR3 is flipped, since those are both unstable according to a la mode.

This surprises me. I am curious to know the reason why we can't make the cavity stable by flipping the PR3 as PR3 was supposed to have more lensing effect than PR2 according to my statement.