40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sun Feb 3 13:20:02 2013, Koji, Summary, General, Hypothesis mode_density_PRC.pdfmode_density_PRC.zip
    Reply  Mon Feb 4 10:45:51 2013, Jamie, Summary, General, rough analysis of aligned PRM-PR2 mode scan scan-labeled.pdf
       Reply  Mon Feb 4 11:10:59 2013, Koji, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
          Reply  Mon Feb 4 19:33:19 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan 3peakdata.png
             Reply  Tue Feb 5 02:04:44 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan PRMPR2scan.png
                Reply  Tue Feb 5 03:16:51 2013, Koji, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
                   Reply  Tue Feb 5 10:09:08 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
                      Reply  Tue Feb 5 11:30:19 2013, Koji, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
                         Reply  Wed Feb 13 01:26:08 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan unbiased.png
       Reply  Wed Feb 6 15:20:55 2013, yuta, Summary, General, FWHM was wrong 
    Reply  Mon Feb 4 15:06:56 2013, Koji, Summary, General, Hypothesis 
    Reply  Mon Feb 4 19:48:32 2013, Jamie, Summary, General, arbcav recalc of PRC with correct ITM transmission mode_density_PRC_2.pdfmode_density_PRC_3.pdf
Message ID: 7998     Entry time: Tue Feb 5 03:16:51 2013     In reply to: 7997     Reply to this: 8000
Author: Koji 
Type: Summary 
Category: General 
Subject: rough analysis of aligned PRM-PR2 mode scan 

0.764 and 0.751 do not give us the stdev of 0.005.

I have never seen any Yokogawa in vicinity.

Quote:

 Measured FWHM was 0.764 MHz and 0.751 MHz. By taking the average, FWHM = 0.757 +/- 0.005 MHz.
 This gives you Finesse = 104 +/- 1, which is OK compared with the expectation.

What I need
 I need better oscilloscope so that we can take longer data (~1 sec) with higher resolution (~0.004 V/count, ~50kHz).
 TDS 3034B can take data only for 10 ksamples, one channel by one!  I prefer Yokogawa DL750 or later.

 