40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sun Feb 3 13:20:02 2013, Koji, Summary, General, Hypothesis mode_density_PRC.pdfmode_density_PRC.zip
    Reply  Mon Feb 4 10:45:51 2013, Jamie, Summary, General, rough analysis of aligned PRM-PR2 mode scan scan-labeled.pdf
       Reply  Mon Feb 4 11:10:59 2013, Koji, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
          Reply  Mon Feb 4 19:33:19 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan 3peakdata.png
             Reply  Tue Feb 5 02:04:44 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan PRMPR2scan.png
                Reply  Tue Feb 5 03:16:51 2013, Koji, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
                   Reply  Tue Feb 5 10:09:08 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
                      Reply  Tue Feb 5 11:30:19 2013, Koji, Summary, General, rough analysis of aligned PRM-PR2 mode scan 
                         Reply  Wed Feb 13 01:26:08 2013, yuta, Summary, General, rough analysis of aligned PRM-PR2 mode scan unbiased.png
       Reply  Wed Feb 6 15:20:55 2013, yuta, Summary, General, FWHM was wrong 
    Reply  Mon Feb 4 15:06:56 2013, Koji, Summary, General, Hypothesis 
    Reply  Mon Feb 4 19:48:32 2013, Jamie, Summary, General, arbcav recalc of PRC with correct ITM transmission mode_density_PRC_2.pdfmode_density_PRC_3.pdf
Message ID: 7991     Entry time: Mon Feb 4 11:10:59 2013     In reply to: 7990     Reply to this: 7994
Author: Koji 
Type: Summary 
Category: General 
Subject: rough analysis of aligned PRM-PR2 mode scan 

The expected finesse is 100ish. How much can we beleive the measured number of 50?
From the number we need to assume PR2 has ~93% reflectivity.
This does not agree with my feeling that the cavity is overcoupled.
Another way is to reduce the reflectivity of the PRM but that is also unlikely from the data sheet.

The scan passed the peak in 4ms according to the fitting.
How do the analog and digital antialiasing filters affect this number?