Fri Jul 25 10:48:13 2008, Koji, Update, General, Abs. Len. Meas. ~ Realignment / beating / PLL trial
|
Mon Jul 28 15:54:04 2008, Koji, Update, General, Abs. Len. Meas. ~ More on the beat / the PLL setup 
|
Thu Jul 31 11:19:30 2008, Koji, Update, General, Abs. Len. Meas. ~ Resonance search trial
|
Sat Aug 2 12:53:43 2008, Koji, Update, General, Abs. Len. Meas. ~ New PLL at the PSL table
|
Mon Aug 4 00:37:58 2008, Koji, Update, General, Abs. Len. Meas. ~ RF PD at the Y end / Manual frequency scan  
|
Mon Aug 4 05:23:57 2008, Koji, Update, General, Abs. Len. Meas. ~ Measurement for Y-arm completed
|
Mon Aug 4 21:48:24 2008, Koji, Update, General, Abs. Len. Meas. ~ Scan for TEM00/01/10  
|
Mon Aug 4 22:31:10 2008, Koji, Update, General, Abs. Len. Meas. ~ Simple Test for TEM01/10 split
|
Tue Aug 5 00:05:57 2008, Koji, Update, General, Abs. Len. Meas. ~ IFR2023A calibration
|
Wed Aug 6 11:10:34 2008, Koji, Update, General, Abs. Len. Meas. ~ analysis of the TEM00 scan 
|
Wed Aug 6 11:43:52 2008, Koji, Update, General, Abs. Len. Meas. ~ analysis of the TEM01 scan
|
Thu Aug 14 21:52:51 2008, Koji, Summary, General, Abs. Len. Meas. ~ summary of my Summer
|
|
Message ID: 795
Entry time: Tue Aug 5 00:05:57 2008
In reply to: 793
Reply to this: 801
|
Author: |
Koji |
Type: |
Update |
Category: |
General |
Subject: |
Abs. Len. Meas. ~ IFR2023A calibration |
|
|
Work log on August 4th
o IFR2023A (Marconi) was calibrated by the SR620 frequency counter which is locked to the GPS signal.
o The frequency of the IFR2023A was scanned from 1MHz to 20MHz with 1MHz interval. The readout of the frequency counter was recorded.
o The linear fit was taken.
f_freq_count = K0 + K1 * f_IFR [Hz]
K0 = 0.00 +/- 0.02
K1 = 0.999999470 +/- 0.000000001
o So, the IFR seems to have -0.5ppm systematic error. |