Fri Jul 25 10:48:13 2008, Koji, Update, General, Abs. Len. Meas. ~ Realignment / beating / PLL trial
|
Mon Jul 28 15:54:04 2008, Koji, Update, General, Abs. Len. Meas. ~ More on the beat / the PLL setup 
|
Thu Jul 31 11:19:30 2008, Koji, Update, General, Abs. Len. Meas. ~ Resonance search trial
|
Sat Aug 2 12:53:43 2008, Koji, Update, General, Abs. Len. Meas. ~ New PLL at the PSL table
|
Mon Aug 4 00:37:58 2008, Koji, Update, General, Abs. Len. Meas. ~ RF PD at the Y end / Manual frequency scan  
|
Mon Aug 4 05:23:57 2008, Koji, Update, General, Abs. Len. Meas. ~ Measurement for Y-arm completed
|
Mon Aug 4 21:48:24 2008, Koji, Update, General, Abs. Len. Meas. ~ Scan for TEM00/01/10  
|
Mon Aug 4 22:31:10 2008, Koji, Update, General, Abs. Len. Meas. ~ Simple Test for TEM01/10 split
|
Tue Aug 5 00:05:57 2008, Koji, Update, General, Abs. Len. Meas. ~ IFR2023A calibration
|
Wed Aug 6 11:10:34 2008, Koji, Update, General, Abs. Len. Meas. ~ analysis of the TEM00 scan 
|
Wed Aug 6 11:43:52 2008, Koji, Update, General, Abs. Len. Meas. ~ analysis of the TEM01 scan
|
Thu Aug 14 21:52:51 2008, Koji, Summary, General, Abs. Len. Meas. ~ summary of my Summer
|
|
Message ID: 793
Entry time: Mon Aug 4 21:48:24 2008
In reply to: 787
Reply to this: 794
795
|
Author: |
Koji |
Type: |
Update |
Category: |
General |
Subject: |
Abs. Len. Meas. ~ Scan for TEM00/01/10 |
|
|
Work log on August 3rd - Part2
o I tried to measure the frequency of the FSRs using TEM00 resonances. Also search of TEM01/TEM10 resonances were tried.
-----------
Measurement for TEM00
o The frequency of the injection beam was scanned from 2MHz to 20MHz using the LabVIEW panel with GPIB. The 1st figure attached below is the result of the scan. Equispaced peaks were found as expected. The interval of the peaks are about 3.89MHz. Each peaks were measured with freq intervals of down to 50Hz. I will analyze the center frequency of the peaks precisely later in order to have a final result.
Measurements for TEM01/TEM10
o The beam injection technique is thought to be useful for measureing the frequency of the higher-order resonances. In order to measure the higher-order resonances the modifications of the experimental setup were applied as below.
1) For TEM10 (the beam like "OO" shape), a razor blade which blocked the horizontal half of the transmitted beam was placed. We needed to disturb half of the beam because the beat between the PSL TEM00 and the injection TEM01 cancels if the PD receives all of the light.
2) The injection beam is slightly misaligned in the horizontal direction in order to enhance the coupling of the injection beam to the cavity TEM01 mode.
3) For TEM01 (the beam like "8" shape), a razor blade cutting the vertical half and the misalignment of the inj beam in the vertical direction are applied.
o The frequency of the injection beam was scaned from 1st FSR of TEM00 in the upward direction. The alignment of the arm cavity was left untouched during the measurement. As shown in the 2nd figure attached below, the resonances were found about 1.19MHz away from the TEM00, but they are separated by about 19kHz(!). This could be split of the degenerated modes which corresponds to the difference of the mirror curvature in two directions! This difference is something like 56 m and 57 m. Can you believe this?
(To be continued to the next entry) |
|
|
|
|
|
|