Message ID: 79     Entry time: Wed Nov 7 14:01:31 2007
Author: waldman 
Type: Omnistructure 
Category: OMC 
Subject: Frequency and Intensity noise 
One of the biggest problems I had using the PZT to lock was excessive noise. I did a little noise hunting and found that the problem was the cable running from the rack to the laser fast input. As a reminder, the laser has a 4 MHz / volt fast input. We require about 300 MHz to go one FSR, so there is a Thorlabs HV box between at the NPRO fast input which takes 0-10 V -> 0-150 V. The 150 V HV range is worth about 600 MHz of NPRO frequency.

OLD SETUP: Single side of DAC differential (10 Vpp) -> 9V in series with 10 kOhm -> 10 kOhm input impedance of Thorlabs HV -> NPRO

We used the single side of the DAC differential because we didn't have a differential receiver. This turned out to be a bad idea because the cable picks up every 60 Hz harmonic known to man kind.

NEW SETUP: Digital conditioning -> DAC differential (digitally limited to 0 - 1 V) -> SR560 in A-B mode gain 10 (0 - 10 V output)-> Thorlabs HV -> NPRO.

This has almost no 60 Hz noise and works much, much better. Moral of the story, ALWAYS USE DIFFERENTIAL SIGNALS DIFFERENTIALLY !

Note that I may be saturating the SR560 with 10 V output, Its spec'd for 10 Vpp output with 1 VDC max input. I don't know whether or not it can push 10 V out....