40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Dec 19 18:40:00 2012, Jenne, Update, IOO, PZT1 removed, TT1 in place 
    Reply  Wed Dec 19 19:28:12 2012, Manasa, Update, IOO, PZT1 removed, TT1 in place 
       Reply  Thu Dec 20 10:11:12 2012, Manasa, Update, IOO, TT1 connections 
          Reply  Thu Dec 20 10:35:27 2012, Steve, Update, IOO, TT1 connections 
Message ID: 7861     Entry time: Thu Dec 20 10:11:12 2012     In reply to: 7858     Reply to this: 7862
Author: Manasa 
Type: Update 
Category: IOO 
Subject: TT1 connections 

[Jamie,Manasa]

We've been trying to figure out the connector for the TTs. Since, we found the cables were plugged in wrong in TT1; when triggered in pitch, the mirror moves in yaw and viceversa.

Referring to the cabling diagram, D1000234-v10, we infer that connectors go as J2 - LR, J3 - UR, J4 - LL, J5 - UL and the connections are made looking at the mirror from behind.

ELOG V3.1.3-