Message ID: 7841     Entry time: Mon Dec 17 19:47:15 2012
Author: rana 
Type: Update 
Category: PEM 
Subject: Seismic StripTool config updated 

I have updated the Seismic Striptool display which is plotted on the wall in the control room. Please take a look and make comments. We should finalize it and not change it anymore.

By having an unchanging display, we can get used to small changes in the seismic environment which disrupt our locking.

  1. y-scale is now linear; the log-scale was suppressing the factor of 2-3 variations which are important to us.
  2. Just as the rainbow does, the colors now go from red to purple to represent the noise from 0.1 - 30 Hz: the red traces are 0.1-0.3 Hz, the green/yellow traces are 0.3-3 Hz, and the blue/purple traces are 3-30 Hz.
  3. This is just showing GUR1. Let's try to keep this seismometer working so that we can have some long term record of the seismicity here. This means don't click off the buttons, disconnect the sensor, reboot the machine, etc. When you do do these things, elog them.
Attachment 1: SeismicRainbow.png  54 kB  | Hide | Hide all
SeismicRainbow.png
ELOG V3.1.3-