Message ID: 7834     Entry time: Fri Dec 14 14:40:31 2012
Author: Riju 
Type: Update 
Category:  
Subject: Photodiode transimpedance 

Photodiode PDA10CF was under test. The RF out signal of AG4395A had been divided by splitter with one output of the splitter going to R channel of the network analyzer and the other to the laser. The splitted laser beams - splitted with beam splitter - fall on two photodiodes - one reference and the other on PDA10CF. The outputs of these two photodiodes go to channel B and A respectively of the network analyzer. The measured transimpedance data had been collected using the GPIB connection.

The result is as follows:

Attachment 1: PDA10CF.pdf  8 kB  | Hide | Hide all
PDA10CF.pdf