Yesterday, I made new mounts for microphones.
 
I glued a microphone on a pedestal. The cables are attached loosely so that its tension does not make any noise.
At the bottom of the mount, I attached the surgical tube forming a ring by double-side tape so that it damps the seismic vibration.
I made 6 mounts and these are all on the AS table now.
I took some data of XARM signal controlled by AS.
My plan is to find/set an upper limit on acoustic coupling noise in AS signal.
The acoustic noise can be estimated by the Wiener filter, but it is not accurate because it may see residual correlation between AS and microphone signals that should be 0 when the data is long enough.
I will find/set an upper limit by the analysis based on Neuman-Pearson criterion, that is analog of a stochastic GW background search.
If I can find the acoustic coupling noise should be below the shot noise, I am happy. If not, some improvements may be needed someday. |