[Rana, Ayaka]

The BS oplev pitch feedback came back.

OL_BSpit.pdf

The problem was that 300^2:0 filter was off. And I turned on all the low pass filters (ELP35), then the oplev servo does not seem to inject big noise into the arms as long as I see the spectra of POY and POX. These low-pass filters will be modified tomorrow so that the acoustic coupling noise is minimized.

BSoplevservo.png