Wed Nov 28 23:18:35 2012, Charles, Update, PEM, Decreased RMS in Seismometers
|
Thu Nov 29 02:43:48 2012, Ayaka, Update, PEM, Decreased RMS in Seismometers  
|
Thu Nov 29 09:58:06 2012, Den, Update, PEM, Decreased RMS in Seismometers
|
Tue Jan 29 14:34:42 2013, Jenne, Update, PEM, Secret Seis filters
|
|
Message ID: 7759
Entry time: Wed Nov 28 23:18:35 2012
Reply to this: 7762
|
Author: |
Charles |
Type: |
Update |
Category: |
PEM |
Subject: |
Decreased RMS in Seismometers |
|
|
The attached plots display RMS noise from various accelerometers and seismometers over the past 90 days. One can see how after the reinstallation of the seismometers in November, RMS from the GUR1Z and GUR1X channels decreases by a factor of about 100 from data in August. Additionally, the RMS over the course of the last 90 days has notably decreased in all instruments. In many cases, the RMS is only the result of inherent electronics noise, rather than from a signal. |
|
|