40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
|
Message ID: 775
Entry time: Thu Jul 31 10:27:17 2008
In reply to: 772
|
Author: |
rana |
Type: |
Update |
Category: |
PSL |
Subject: |
PMC Scan Graphs |
|
|
Quote: | Graphs of the PMC scan data that I got earlier today.
|
On the UNIX computers, one can use 'convert' to change these to PNG. A DC offset should be added to the transmitted
light so that the scan can be plotted with a log y-scale. And, of course, Acrobat can be used to make it into a
single PDF file.
The PMC scan always has this distortion and so the input power has to be decreased to a few mW to reduce the
thermal expansion effect; the expansion coefficient for SiO2 is ~5 x 10^-7 / K and we're worried about nm level
expansions. |