Entry  Wed Oct 17 08:04:31 2012, Steve, Update, VAC, Roughing pumps on for oil change 
    Reply  Tue Oct 23 16:18:03 2012, Steve, Update, VAC, Roughing pumps on for oil change oilchange.jpg
Message ID: 7597     Entry time: Tue Oct 23 16:18:03 2012     In reply to: 7564
Author: Steve 
Type: Update 
Category: VAC 
Subject: Roughing pumps on for oil change 

Leybold D30AQuote:

PR1, PR2 and RP3 turned on for warming up for oil change. Oil changed with 3.2L of MVT-19  fluid in each. This substitute for HE-175 will be used next time.

To finish this job tomorrow: 1, check oil creeping upstream  2, change air filter of air purge if pressure drops <350 mTorr  3, measure venting time of pump

 . Leybold D30A manual is here. Exhaust filter traps were drained. No oil creeping was found. The pump venting time is < 1 minute.

PR-2 needs new secu-valve. PR1 & 2 are in excellent condition.

Attachment 1: oilchange.jpg  2.298 MB  Uploaded Wed Oct 24 08:37:14 2012  | Hide | Hide all