[Koji / Kiwamu]
We have realigned the interferometer except the incident beam.
The REFL beam is not coming out from the chamber and is likely hitting the holder of a mirror in the OMC chamber. 
So we need to open the chamber again before trying to lock the recycled interferometers at some point.
--- What we did
- Ran the MC decenter script to check the spot positions.
- MC3 YAW gave a - 5mm offset with an error of about the same level.
- We didn't believe in this dither measurement.
- Checked the IP-POS and IP-ANG trends.
- The trends looked stable over 10 days (with a 24 hours drift).
- So we decided not to touch the MC suspensions.
- Tried aligning PRM
- Found that the beam on the REFL path was a fake beam
- The position of this beam was not sensitive to the alignment of PRM or ITMs.
- So certainly this is not the REFL beam.
- The power of this unknown beam is about 7.8 mW
- Let the PRM reflection beam go through the Faraday
- This was done by looking at the hole of the Faraday though a view port of the IOO chamber with an IR viewer.
- Aligned the rest of the interferometer (not including ETMs)
- We used the aligned PRM as the alignment reference
- Aligned ITMY such that the ITMY reflection overlaps with the PRM beam at the AS port.
- Aligned the BS and SRM such that their associated beam overlap at the AS port
- Aligned ITMX in the same way.
- Note that the beam axis, defined by the BS, ITMX and SRM, was not determined by this process. So we need to align it using the y-arm as a reference at some point.
- After the alignment, the beam at the AS port still doesn't look clipped. Which is good.
---- things to be fixed
- Align the steering mirrors in the faraday rejected beam path (requires vent)
- SRM oplev (this is out of the QPD range)
- ITMX oplev (out of the range too) |