Entry  Tue Oct 2 07:45:49 2012, Steve, Update, SUS, MC2 & PRM sus damping restored 
    Reply  Wed Oct 3 07:41:23 2012, Steve, Update, SUS, MC2 sus damping restored 
       Reply  Wed Oct 3 23:24:05 2012, Koji, Update, SUS, MC2 sus damping restored 
       Reply  Sun Oct 7 15:05:42 2012, Den, Update, SUS, MC2 sus damping restored mc2.png
Message ID: 7496     Entry time: Sun Oct 7 15:05:42 2012     In reply to: 7466
Author: Den 
Type: Update 
Category: SUS 
Subject: MC2 sus damping restored 

Quote:

This is the third morning in a row that the MC2 was tripped.

 MC2 was tripped again. I think the answer is that watchdog's critical value was too small C1:SUS-MC2_PD_MAX_VAR = 10, so seismic could trip MC2. I've changed the value to 100.

mc2.png