Message ID: 7465     Entry time: Tue Oct 2 16:32:43 2012     In reply to: 7463     Reply to this: 7475
Author: Jenne 
Type: Update 
Category: IOO 
Subject: PZT diagnosis 

[Koji, Jenne]

Jamie and I pulled the whole PZT driver for both PZT1 and PZT2. 

Koji and I found that each HV power supply (the left-most module) has 2 fuses.  Both HV supplies (PZT1 and PZT2) have one blown fuse.  The "T2L250A" measures low resistance for both HV supplies, but the "T250mAL250V" measures Open for both HV supplies.

I have ordered 10 pieces of each kind of fuse, Next Day shipping, from DigiKey.