40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Tue Sep 18 20:32:42 2012, Manasa, Configuration, PSL, AOM installation
|
Wed Sep 19 11:39:37 2012, Manasa, Configuration, PSL, AOM installation
|
Wed Sep 19 15:41:27 2012, Manasa, Configuration, PSL, AOM installation
|
Wed Sep 19 23:17:25 2012, rana, Configuration, PSL, AOM installation
|
Thu Sep 20 01:29:04 2012, Manasa, Configuration, PSL, AOM installation
|
Fri Sep 21 12:12:56 2012, Manasa, Configuration, PSL, AOM installation
|
Tue Oct 2 16:15:22 2012, Manasa, Configuration, PSL, AOM installation
|
Wed Oct 3 16:52:16 2012, Manasa, Configuration, PSL, AOM installation
|
Wed Oct 3 23:36:54 2012, Koji, Configuration, PSL, AOM installation
|
Thu Oct 4 17:54:59 2012, Manasa, Configuration, PSL, AOM installation
|
Thu Oct 4 18:48:04 2012, janosch, Configuration, PSL, AOM installation
|
Thu Oct 4 20:57:43 2012, Manasa, Configuration, PSL, AOM installation
|
Thu Oct 4 22:16:28 2012, Koji, Configuration, PSL, AOM installation
|
Fri Oct 5 18:08:17 2012, Manasa, Configuration, PSL, AOM installation
|
|
Message ID: 7425
Entry time: Fri Sep 21 12:12:56 2012
In reply to: 7416
Reply to this: 7464
|
Author: |
Manasa |
Type: |
Configuration |
Category: |
PSL |
Subject: |
AOM installation |
|
|
{Jan, Manasa}
We installed the AOM driver back on the PSL table this morning. To calibrate the AOM RF output we connected a 1V dc to the modulation input of the driver and we are convinced with the setup.
Before we direct the rf signal to the AOM, in order to check its diffraction efficiency, we would like to setup an rf PD at the AOM output. We think we have place for a filter and PD after the AOM (replacing a beam dump) and would like to confirm the position before we actually install them. The layout is the picture below showing sweet spots for the new pd to sit. If you think it may disturb the system in any way, let us know!

|