Fri Aug 24 22:01:15 2012, Jenne, Update, IOO, MC spot position - close
|
Sun Aug 26 12:26:44 2012, Jenne, Update, IOO, MC spot position - not done yet
|
Mon Aug 27 01:05:36 2012, Jenne, Update, IOO, MC spot position - callin' it quits
|
Mon Aug 27 12:03:54 2012, Koji, Update, IOO, MC spot position - callin' it quits
|
Mon Aug 27 18:32:48 2012, Jenne, Update, IOO, MC spot position - Jenne is stupid
|
Mon Aug 27 18:59:24 2012, jamie, Update, IOO, MC ASC screen was confusing - Jenne is not stupid 
|
Mon Aug 27 23:52:59 2012, Jenne, Update, IOO, MC Spots centered
|
|
Message ID: 7289
Entry time: Mon Aug 27 18:59:24 2012
In reply to: 7288
Reply to this: 7290
|
Author: |
jamie |
Type: |
Update |
Category: |
IOO |
Subject: |
MC ASC screen was confusing - Jenne is not stupid |
|
|
Quote: |
We have figured out that some of these measurements, those with the WFS off, were also not allowing the dither lines through, so no dither, so no actual measurement.
Jamie is fixing up the model so we can force the WFS to stay off, but allow the dither lines to go through. He'll elog things later.
|
In the c1ioo model there were filter modules at the output of the WFS output matrix, right before going to the MC SUS ASCs but right after the dither line inputs, that were not exposed in the C1IOO_WFS_OVERVIEW screen (bad!). I switched the order of these modules and the dither sums, so these output filters are now before the dither inputs. This will allow us to turn off all the WFS feedback while still allowing the dither lines.
I updated the medm screens as well (see attached images). |
|
|
|
|