I estimated the transfer function of the seismic stacks using a rough model I made based on the LIGO document LIGO T000058 -00. I used a Q of 3.3 for the viton springs, and resonant frequencies of 2.3, 7.5, 15, and 22 Hz (measured in that document for the horizontal motion). I multiplied the simple mass-spring transfer function four times for each layer of metal/spring, with the respective resonant frequency for each. The pendulum suspending the test masses has a resonant frequency of 0.74 and a Q of 3, according to the same document.
 
When I multiply the net transfer function (pendulum included, the green line above) by the differential motion of the x arm that I measured in eLog 7186, I find the differential motion of the test mass (NOTE: I converted the differential motion to displacement by multiplying by (1/2*pi*f)).
 
It agrees within an order of magnitude to the seismic wall from the displacement noise spectrum hanging above the control room computers.
Finally, I looked at how the geophone and accelerometer noise spectra looked compared to the ground differential motion (any STACIS sensor signal will also be multiplied by the stack/pendulum transfer function, so I'm comparing to the differential motion before it goes through the chamber). Below about 1 Hz, it is clear from the plot below that the STACIS could never be of any benefit, even with accelerometers rather than geophones as the feedback sensors.
 
|