40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Aug 15 17:17:58 2012, Manasa, Jan, Update, IOO, Ringdown measurements Ringdown_815.jpg
    Reply  Wed Aug 15 20:15:51 2012, Jan, Update, IOO, Ringdown measurements 
       Reply  Wed Aug 15 20:53:48 2012, Manasa, Update, IOO, Ringdown measurements 
Message ID: 7199     Entry time: Wed Aug 15 20:15:51 2012     In reply to: 7196     Reply to this: 7200
Author: Jan 
Type: Update 
Category: IOO 
Subject: Ringdown measurements 

Quote:

While I thought that the bumps observed at the end of the ringdown might be because of the cavity trying to lock itself, Jan commented that they have always existed in these measurements and their source is not known yet.

What I meant to say was that in all ringdown measurements that we observed today, the bumps were consistently part the ringdown, and that I have no explanation for the bumps. It should also be mentioned that fitting the bumpy part of the ringdown instead (we used the clean first 10us), the ringdown time is about twice as high. In either case, the ringdown time is significantly smaller than we have seen in documents about previous measurements.

We (basically I) also made one error when producing the plots. The yaxis label of the semi-logarithmic plot should be log(...), not log10(...).