40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 7186     Entry time: Wed Aug 15 01:14:19 2012
Author: Yaakov 
Type: Update 
Category: PEM 
Subject: Differential Motion of X and Y Arm 

Den and I measured the differential motion of the x and y arms using Guralp 1 at the end of the y arm, Guralp 2 at the beamsplitter, and the Streckeisen at the end of the x arm.

I calibrated the Streckeisen to the Guralp by calculating the relative gain of the seismometer signals at the microseism. The Guralp 1-y amplitude was 1.0237 times Guralp 2-y and Guralp 2-x was 38.54 times STS-x. The Guralp calibration (to go from counts to meters) I used was 0.61/1000/800/80/(2*pi*f) m/count.

The differential motion should keep decreasing at low frequencies because the ground will move together at such large wavelengths. It goes up because the seismometer noise begins to dominate at low frequencies (below about 0.5 Hz). Another possible error source could be that the seismometers are not perfectly aligned along the arm.

diff_motion_x_arm.pngdiff_motion_y_arm.png

Attachment 1: diff_motion_x_arm.png  38 kB  | Hide | Hide all
diff_motion_x_arm.png
Attachment 2: diff_motion_y_arm.png  38 kB  | Hide | Hide all
diff_motion_y_arm.png
ELOG V3.1.3-