Entry  Thu Jul 5 13:38:05 2012, yuta, Summary, Locking, cavity g-factor from mode scan 
    Reply  Sun Jul 8 00:27:54 2012, yuta, Summary, Locking, calibrating phase tracking mode scan data 
       Reply  Sun Jul 8 00:58:08 2012, Koji, Summary, Locking, calibrating phase tracking mode scan data 
Message ID: 6939     Entry time: Sun Jul 8 00:58:08 2012     In reply to: 6938
Author: Koji 
Type: Summary 
Category: Locking 
Subject: calibrating phase tracking mode scan data 

Quote:

FSR for X/Y arm are 3.97 +/- 0.03 MHz and 3.96 +/- 0.02 MHz respectively. This means X/Y arm lengths are 37.6 +/- 0.3 m and 37.9 +/- 0.2 m respectively.

These aren't so bad. (Look at this entry)

And interestingly the ETM curvatures are closer to ATF measurements than Coastline's measurement. (Look at wiki)