40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sat Jun 23 02:29:18 2012, yuta, Update, Green Locking, X arm mode scan results XarmScan20120623.pngfine8FSRscanXarm.png
    Reply  Wed Jun 27 03:43:52 2012, yuta, Summary, IOO, how to improve mode matching to arms modematchMCtoARM_design.pngmodematchMCtoARM_MMT1MMT2longer.png
       Reply  Wed Jun 27 10:27:09 2012, rana, Summary, IOO, how to improve mode matching to arms 
Message ID: 6877     Entry time: Wed Jun 27 10:27:09 2012     In reply to: 6876
Author: rana 
Type: Summary 
Category: IOO 
Subject: how to improve mode matching to arms 

The MC waist is correct as is the arm RoCs. Most likely the error is in the telescope length or its distance from the MC. Jenne probably has all the numbers and can give us a surface plot showing how the MM degrades as a function of those two parameters.