Message ID: 6723     Entry time: Thu May 31 01:20:41 2012
Author: Jenne 
Type: Update 
Category: ASS 
Subject: ASS filter outputs are non-zero with no input 

I was looking a little at ASS, while Yuta was doing some Green transmitted DC PD work, and I find that the output of some filters is totally insane with no deliberate input or excitation signals.

Note in the figure that the filter (which is a 2nd order butter bandpass in the C1:ASS-LOCKIN29_SIG filter bank) is ringing a lot - this needs fixing.  But, more disconcertingly, sometimes (not every time) the arm flashes, the input to the filter bank gets a ~1 sample long spike that is ~9,000,000 counts.  9 million is a lot of counts.  This is then making the filter go crazy.

Any ideas on how this can happen, and how we can stop / fix it?  It's certainly a CDS issue, but I'm not sure where or how.