Suresh noted that I never wrote down the waist positions of the beam propagated through the MMT (based on where we think it is from ruler-based measurements). This uses the MC waist measured beam as the starting point, and just propagates through the MMT, out to the IPPOS port.
Yq:
Properties:
q: 2.2488 +23.8777i
lambda: 1.0640e-06
waistSize: 0.0028 (at z = 13.2676 meters)
waistZ: 2.2488 (relative to z = 13.2676 meters)
divergenceAngle: 1.1910e-04
radiusOfCurvature: 255.7849
beamWidth: 0.0029
rayleighRange: 23.8777
So, to sum up, the vertical beam waist is 2.8438 mm at 11.0188 meters from the MC waist.
Xq:
Properties:
q: 5.1953 +24.7342i
lambda: 1.0640e-06
waistSize: 0.0029 (at z = 13.2676 meters)
waistZ: 5.1953 (relative to z = 13.2676 meters)
divergenceAngle: 1.1702e-04
radiusOfCurvature: 122.9525
beamWidth: 0.0030
rayleighRange: 24.7342
So, to sum up, the horizontal beam waist is 2.8943 mm at 8.0723 meters from the MC waist.
In pictorial form,

|