Keiko, Rana, Suresh
Related to the beam profile of IPPOS today, we tried to measure the beam size at the ETMY point in order to estimate the input beam mode. We measured the beam size hitting at the suspension frame by a camera image, with two situations to see two "z" for beam profile.
(1) Input beam is slightly misaligned and the injected beam hits the end mirror frame. Assuming z=0 at the input mirror, this should be z=40m.
(2) Input beam hits the centre of the end mirror, and ITMY is slightly misaligned and the beam hits the end mirror frame after the one-round trip. Assuming z=0 at the ITM, this position should be z=120m.

The injected beam at the end point and the one round trip ligt at the end point should be the same size, if the input mode matches to the cavity mode. You can see if your injected light is good for the cavity or not. We compared and assumed the above two beam sizes by looking at the photo of the beam spot.
(1) (2) 
Assuming the zoom factor difference by the part below the beam (shown with allow in the photos. Arbitrary unit.), the beam in (2) is smaller than expected (roughly 40%?).
However this is a very rough estimation of the beam sizes! It is difficult to assume the beam size shown on the photos! It looks smaller only because the power of (2) is smaller than of (1). I don't think we can say anything from this rough estimation. One may be able to estimate better with CCD camera instead of this normal camera.
|