Entry  Wed Mar 21 10:31:28 2012, steve, Update, SAFETY, AC power to ETMY turned off 
    Reply  Wed Mar 21 11:32:47 2012, steve, Update, SAFETY, AC power to ETMX turned off 
       Reply  Wed Mar 21 19:12:27 2012, steve, Update, Computers, AC power back on both ends 
          Reply  Thu Mar 22 16:45:06 2012, kiwamu, Update, CDS, c1scx and c1scy not properly running 
             Reply  Thu Mar 22 17:41:15 2012, suresh, Update, CDS, c1scx and c1scy not properly running 
                Reply  Thu Mar 22 23:43:56 2012, Koji, Update, CDS, c1scx and c1scy not properly running 
Message ID: 6438     Entry time: Thu Mar 22 17:41:15 2012     In reply to: 6436     Reply to this: 6439
Author: suresh 
Type: Update 
Category: CDS 
Subject: c1scx and c1scy not properly running 

Quote:

It seems that neither c1scx nor c1scy is working properly as their ADC counts are showing digital-zeros.

Quote from #6434

The power was turned back on at 4pm It took some time for Suresh to restart the computers. We have damping but things are not perfect yet. Auto BURTH did not work well.

 When Steve and I restarted the c1iscex and c1iscey computers after the power shutdown, the models within them did not start-up automatically.  I had to start them manually from a terminal in the control room. 

I also tried rebooting the FB a couple of times.  Did not make any difference.

Manually starting the c1x05, c1scy and c1x01, c1scx models (with the Burt Restore button ON) did not resolve the issue of zeros in the epics screens.  though it did re-establish timing. 