Entry  Thu Feb 16 21:13:07 2012, Suresh, Update, Electronics, REFL165 repair: PD replaced, DC response checked with a torch light PC_30641_old.jpgPD_30641_new.jpgIMG_0548.JPG
    Reply  Tue Feb 28 16:31:08 2012, Suresh, Update, Electronics, REFL165 repair: Characterization REFL165_Characterization.pdfREFL165_response_shift.pdf
       Reply  Wed Feb 29 00:22:35 2012, Suresh, Update, Electronics, REFL165 repair: Installed on the AS table 
       Reply  Wed Feb 29 01:14:40 2012, Suresh, Update, Electronics, REFL165 repair: Characterization 
Message ID: 6337     Entry time: Wed Feb 29 00:22:35 2012     In reply to: 6333
Author: Suresh 
Type: Update 
Category: Electronics 
Subject: REFL165 repair: Installed on the AS table 

1) The REFL165 has been replaced onto the AS table.

2) When the PD interface cable is attached the PD shows a DC out put of 6mV and does not respond to a flash light.  I changed the PD interface port in the LSC rack by swapping the other end of the cable with an unused (Unidentified PD) interface cable,  The PD is working fine after that.   There could be a problem with some binary switch state on the PD interface where the REFL165 cable was plugged in earlier.

 