40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Feb 6 17:06:11 2012, steve, Update, SUS, oplev transferfunction of PIT: ETMY & ETMX ETMX_PIT.pngETMY_PIT.png
    Reply  Tue Feb 7 16:30:46 2012, steve, Update, SUS, ALL oplev PIT transfer functions 7x
       Reply  Thu Feb 9 16:46:02 2012, steve, Update, SUS, ALL oplev YAW transfer functions 7x
          Reply  Thu Feb 23 10:53:54 2012, steve, Update, SUS, oplev PIT transfer functions oplPITtrans.pdf
             Reply  Mon Feb 27 14:35:22 2012, steve, Update, SUS, oplev YAW transfer functions oplYAWtransf.pdf
Message ID: 6323     Entry time: Mon Feb 27 14:35:22 2012     In reply to: 6309
Author: steve 
Type: Update 
Category: SUS 
Subject: oplev YAW transfer functions 

 The BS and the PRM have 3.3 Hz resonant gain filters that kill the phase margins.

 

Attachment 1: oplYAWtransf.pdf  84 kB  | Hide | Hide all
oplYAWtransf.pdf
ELOG V3.1.3-