Entry  Fri Feb 24 03:58:13 2012, kiwamu, Update, LSC, Y arm + PRMI part II 
    Reply  Fri Feb 24 15:01:31 2012, kiwamu, Update, LSC, Y arm + PRMI part II time_series.png
       Reply  Fri Feb 24 19:18:28 2012, kiwamu, Update, LSC, Y arm + PRMI : how they should look like expected_time_series.png
    Reply  Fri Feb 24 18:37:13 2012, rana, Update, LSC, Y arm + PRMI part II 
Message ID: 6315     Entry time: Fri Feb 24 18:37:13 2012     In reply to: 6310
Author: rana 
Type: Update 
Category: LSC 
Subject: Y arm + PRMI part II 

Quote:

I tried the Yarm + PRMI configuration again.

The PRMI part was locked, but it didn't stay locked during the Y arm was brought to the resonance point.

 

 Isn't the point that the 11 and 55 MHz signals have the carrier effect, but the 3f signals are better?