Entry  Wed Nov 23 18:28:09 2011, Koji, Update, IOO, RFAMPD channels / EOM monitor channels added to DAQ 
    Reply  Thu Nov 24 05:10:00 2011, Koji, Update, IOO, RFAMPD channels / EOM monitor channels added to DAQ 
       Reply  Thu Nov 24 10:27:07 2011, Jenne, Update, IOO, RFAMPD channels / EOM monitor channels added to DAQ EOM_temp_sense_heater_drive_schematic_withMONs.png
          Reply  Thu Nov 24 12:45:12 2011, Zach, Update, IOO, RFAMPD channels / EOM monitor channels added to DAQ 
             Reply  Thu Nov 24 13:54:31 2011, Koji, Update, IOO, RFAMPD channels / EOM monitor channels added to DAQ 
Message ID: 5991     Entry time: Wed Nov 23 18:28:09 2011     Reply to this: 5995
Author: Koji 
Type: Update 
Category: IOO 
Subject: RFAMPD channels / EOM monitor channels added to DAQ 

The following channels have been registered in c1iool0 database, and are now recorded by FB

C1:IOO-RFAMPD_11MHZ
C1:IOO-RFAMPD_29_5MHZ
C1:IOO-RFAMPD_55MHZ
C1:IOO-RFAMPD_DCMON
C1:IOO-EOM_TEMPMON
C1:IOO-EOM_HEATER_DRIVEMON


PROCEDURE

1) The EPICS database file has been edited to rename/add some channels

/cvs/cds/caltech/target/c1iool0/ioo.db

REMOVED
#grecord(ao,"C1:IOO-RFAMPD_VC")
#grecord(ai,"C1:IOO-RFAMPD_TEMP")
#grecord(ai,"C1:IOO-RFAMPD_DCMON")
#grecord(bo,"C1:IOO-RFAMPD_BIAS_ENABLE")
#grecord(bi,"C1:IOO-RFAMPD_BIAS_STATUS")
#grecord(calc, "C1:IOO-RFAMPD_33MHZ_CAL")
#grecord(calc, "C1:IOO-RFAMPD_133MHZ_CAL")
#grecord(calc, "C1:IOO-RFAMPD_166MHZ_CAL")
#grecord(calc, "C1:IOO-RFAMPD_199MHZ_CAL")

ADDED/EDITED
grecord(ai,"C1:IOO-RFAMPD_11MHZ")
        field(DTYP,"VMIVME-3113")                                              
        field(INP,"#C1 S25 @")
...

grecord(ai,"C1:IOO-RFAMPD_29_5MHZ")
        field(DTYP,"VMIVME-3113")                                              
        field(INP,"#C1 S26 @")

...
grecord(ai,"C1:IOO-RFAMPD_55MHZ")
        field(DTYP,"VMIVME-3113")                                              
        field(INP,"#C1 S27 @")

...
grecord(ai,"C1:IOO-RFAMPD_DCMON")
        field(DTYP,"VMIVME-3113")                                              
        field(INP,"#C1 S28 @")

...
grecord(ai,"C1:IOO-EOM_TEMPMON")
                                                
        field(DTYP,"VMIVME-3113")                                               
        field(INP,"#C1 S29 @")

...
grecord(ai,"C1:IOO-EOM_HEATER_DRIVEMON")

        field(DTYP,"VMIVME-3113")                                              
        field(INP,"#C1 S30 @")

2) The channels have been added to the frame builder database

/cvs/cds/rtcds/caltech/c1/chans/daq/C0EDCU.ini

[C1:IOO-RFAMPD_11MHZ]
[C1:IOO-RFAMPD_29_5MHZ]
[C1:IOO-RFAMPD_55MHZ]
[C1:IOO-RFAMPD_DCMON]
[C1:IOO-EOM_TEMPMON]
[C1:IOO-EOM_HEATER_DRIVEMON]

Note that this C0EDCU.ini is the file that has been registered in

/cvs/cds/rtcds/caltech/c1/target/fb/master

3) burt restore request files were updated

RFAM related settings were removed as they don't exist anymore.

/cvs/cds/caltech/target/c1iool0/autoBurt.req
/cvs/cds/caltech/target/c1iool0/
saverestore.req

4) c1iool0 were rebooted. Framebuilder restarted. c1iool0 were burtrestored.