[Den, Mirko]
We looked some more into the the OSEM signals and their coherence to the seismometer signals.
We were able to verify that the coherence OSEM sensor <-> seismometer signal goes down with increasing the OSEM gain. This seems to indicate that the OSEM FB add noise to the distance mirror <-> frame. We injected some noise into the OSEMs to see how the coherence behaves.
MC2 SUSPOS, 0.1Hz - 0.8Hz, 3mins each
Inj. amplitude Time(UTC) Note
- 21:35 Free swinging
- 21:42 Big LF OSEM gain
- 21:48 Small LF OSEM gain
150 21:56 -"-
300 22:00 -"-
900 22:05 -"-
Free swinging:

High OSEM gain:

Low OSEM gain:




We left the filters that lower the OSEM gain below 0.3Hz on. |