Message ID: 5920     Entry time: Thu Nov 17 03:46:52 2011
Author: kiwamu 
Type: Update 
Category: Green Locking 
Subject: PSL doubling had been diabled 

I found that the temperature controller of the PSL doubling oven had been disabled.

Because of that I took a little bit long time to recover the beat-note.
I have no idea why its been disabled.
I turned it on to make the PSL green beam bright,
Also the I-parameter of the PID temperature control was too big
and because of that a big overshoot in the temperature happened (overshoot of ~ 5 deg !).
So I decreased the I-parameter from 175 to 85 (250 is the maximum).
Now the intensity of the green light seems reasonably bright and stable.