[Micro, Den]

Analyzing coherence of seismic noise and mode cleaner length we've figured out that at some days the coherence below 1 Hz is still present. For example, at Nov 13 we can see some coherence compared to most other dates when we are not able to see coherence as shown on the figure. On the top plot - psd of MC_L and GUR1_X at Nov 13 (red and blue) and Nov 15 (black and cyan). On the bottom plot is presented coherence between MC_L and GUR1_X on Nov 13 (red) and Nov 15 (black)

datespsd.jpg

datescoh.jpg

We can divide the psd plot for 2 parts - below 1 Hz and above 1 Hz. Above 1 Hz seismic noise on Nov 15 (cyan) was higher then on Nov 13 (blue) and correspondently MC_L at that region was higher on Nov 15. Below 1 Hz seismic noise was higher on Nov 13 but MC_L is still lower that on Nov 15. That is surprising. From the coherence plot we can say that once we have some more seismic noise than usually, we immediately see coherence.

Because of this we wanted to find out the level of the X noise that makes seismic noise invisible. We injected seismic noise by doing smooth physical exercises near MC_2 (1.5 m and 3 m apart). The MC_2 was in lock during the experiment.

injectionpsd.jpg

injectioncoh.jpg

In the coherence plot we can see that coherence between GUR1_X and MC_L increased with noise injection. The highest coherenced we obtained sittind down and standing up smoothly near MC_2 at distance 1.5 m. We did not want to come clother and break the lock. This measurement tells us that the X noise is approximately 3-4 times higher than seismic noise in the range 0.1 - 1 Hz. That means that it is approximately 1e-6 - 1e-8 m/sqrt(Hz) in this region. This noise goes down at frequencies from 2 Hz and not seen because of seismic noise. Actually, seismic noise can be filtered out with the Wiener filter and then we'll see the spectrum of X noise.

We now try to figure out the method to estimate the contribution of OSEM noise to the X noise.