Message ID: 5882     Entry time: Sat Nov 12 02:46:13 2011
Author: Den 
Type: Update 
Category: Adaptive Filtering 
Subject: stacks and ground 

We measured the coherence between the seismometer near the MC2 stack and accelerometers on the vacuum tank where MC2 is. Because accelerometers produce small signals at low frequencies, which are comparable with adc noise, we  amplified the accelerometer signal by a factor of 20. We could not do it more because though adc has 40 V range, the black box that follows the channel sockets can transmit only 2.5 V max amplitude signal. Probably, this was done because old adc accepted 2 V max amplitude.

ground_stack.jpg

ground_stack_coherence.jpg


We were able to found some coherence at 0.1-1 Hz though the accelerometer signal is rather noisy. So to consider stack as a noise source is still possible. This measurement should be better done with two seismometers, one on the floor, the other on the stack. From the figure we can also see that tilt affects the x and y seismometer signals from 0.1 Hz. Green line (z-component) is much lower that red and blue lines (x and y). Tilt affects on horisontal axes of the seismometer much more than on vertical.

What we also think about is that at low frequencies mirrors start to move approximately the same and seismometers can help us to figure out small reletive displacement of the mirrors which form the MC length. We can estimate the critical frequency by presenting the ground motion as interference of surface waves with different velocities and amplitudes. For only 1 wave we have for the relation of MC length to the seismometer read out  ~sin (2*pi*f/v*L). f - frequency of the wave, v -speed, L - length between the mirrors. We can see that below 1 Hz we have ~sin (f/2). At this point seismometer signal could lose coherence with MC length signal. We could try to subtract seismometer signals from corresponding axes, but gur1 and sts1 has different calibrations. Moreover, the noise floor of the seismometers might not allow us to measure the differential signal. We'll try to simulate this scenario and find seismometer calibration or measure it. We are basicly interested only in the ratio of calibraion fucntions of 2 seismometers.

ELOG V3.1.3-