Message ID: 5867     Entry time: Thu Nov 10 22:00:38 2011     In reply to: 5864
Author: Mirko 
Type: Update 
Category: Adaptive Filtering 
Subject: Looking into MC_F & PSL misalignment 

Quote:

 [Den, Mirko]

While doing the things below we accidentally misaligned the PSL laser. Thanks to Suresh and Jenne for realigning!!

There are a lot of strange features in MC_F (see for example http://nodus.ligo.caltech.edu:8080/40m/5738 )
To get some better understanding of the signals in the control loop we looked some more into what happens to the MC feedback signal after it exits the MC servo board (D040180 see DCC).

10112011076.jpg
The MC_F signal is actually the servo signal: http://nodus.ligo.caltech.edu:8080/40m/5695
The Thorlabs temperature controller is actually used in the PZT path!

 We measured the LP filter in the PZT path (that is kind of mislabeled as temp.)

10112011073.jpg

 

 

We looked into the signal from the MC servo board at different position at the PSL table.


We looked into the FB going into the temp. and PZT parts of the FB.
Temp.:
Screenshot.png

PZT:

Screenshot-1.png
We also looked at the signal in just in front of the FSS box No idea why the elog doesn't preview these pdfs.

Signal_at_MC_servo_board_and_PSL_table_2_1.pdf

Signal_at_MC_servo_board_and_PSL_table_2_2.pdf

Signal_at_MC_servo_board_and_PSL_table_2_3.pdf

Signal_at_MC_servo_board_and_PSL_table_2_4.pdf
Lots of extra noise there. We will check out where that comes from.