The alarm handler has been flipping out saying that the LO input of the MC's demod board is too low, so at Rana's suggestion, Eric and I measured the response of the LO input. We used an SR345 function generator at 29.485MHz and several different amplitudes to make a table. The demod board should see an input from the LO between 0-2dBm. When I measured what was going into the LO input from the 29.5MHz delay line phase shifter, the LO input was seeing 4dBm. I'm going to put a 3dB attenuator between the phase shifter and the demod board.
Also, now that we have this table of response values, I'm going to change the settings of the alarm handler to something more reasonable.
Amplitude of 29.485MHz input sine wave [dBm] | Value of channel C1:IOO-MC_DEMOD_LO
-------------------------------------------- | -----------------------------------
-10 | -0.000449867
-8 | -0.000449867
-6 | -0.000449867
-4 | 0.000384331
-2 | 0.00526733
0 | 0.0199163
2 | 0.0492143
4 | 0.0931613
6 | 0.161523
8 | 0.229885
10 | 0.293364 |