Subject: PDH box #17 modified, too 

I also modified the Y-Arm PDH box itself slightly. Previously, there was a flying 10k resistor from the SWEEP input to TP2. I don't see the point of this, so I moved it from TP2 across R19 (to the same point where it is on the gyro PDH boxes) to allow for excitation signals to be injected with the loop closed (i.e., with the SWEEP switch off). This is useful for OLTF measurements.