Message ID: 5635     Entry time: Fri Oct 7 22:48:26 2011
Author: Suresh 
Type: Update 
Category: IOO 
Subject: MC2 Trans QPD spot size and incident power decreased 

After centering the spot on the MC2, I started to adjust the spot position on MC_TRANS_QPD to center the beam on it.  I noticed that the spot size was about 3 to 4mm dia. because the 200mm lens was too close to the QPD.  I moved it back and decreased the spot size to about 1mm and the sensitivity to spot position increased.  However, Koji noted that the QPD sectors were near saturation, so I put in a ND=0.3 filter to reduce the incident power on the QPD.

At optimal alignment the current QPD_SUM is around 25k to 26k counts (factor of 2 down). Eventually the gain of the QPD ckts have to be reduced to prevent saturation, for the moment this is temporary fix.

The MC_TRANS_SUM trigger for MC autolocker is working fine no further change was required.