40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 5615     Entry time: Tue Oct 4 16:10:45 2011
Author: Suresh 
Type: Update 
Category: Computer Scripts / Programs 
Subject: MC was not damping 

The MC was not damping earlier this morning around 11:45AM.  The reason was that the INMATRIX on all the three MC1, 2 and 3 were zero.   This was seen earlier when autoburt did not restore this.

This got fixed when I Burt-Restored c1mcsepics.snap

In the afternoon we had to restore c1mcs again to restore the MC_TRANS channels because its INMATRIX was also zero.

Can we do something to make sure this gets done in the autoburt properly?

ELOG V3.1.3-