I created 3 kinds of LSC matrices, PRMI condition with carrier resonant in PRC, PRMI condition with SB resonant in PRC, and DRMI with SB resonant in PRC. The matrices are with AS55 and REFL11 which are used for locking right now. The signal numbers are written in log10, and the dem phases are shown in degrees.
From CR reso PRMI to SB reso PRMI, demodulation phases change ----
PRMI - Carrier resonant in PRC
PRCL MICH SRCL
REFL11 |
7.7079 |
2.9578 |
0 |
REFL33 |
5.2054 |
3.2161 |
0 |
REFL55 |
7.7082 |
2.9584 |
0 |
REFL165 |
3.9294 |
2.5317 |
0 |
AS11 |
1.0324 |
3.5589 |
0 |
AS33 |
1.0286 |
1.6028 |
0 |
AS55 |
1.1708 |
4.2588 |
0 |
AS165 |
1.1241 |
0.9352 |
0 |
POP11 |
2.8015 |
-1.3331 |
0 |
POP33 |
0.2989 |
-1.6806 |
0 |
POP55 |
2.8017 |
-0.6493 |
0 |
POP165 |
-0.9769 |
-2.3708 |
0 |
POX11 |
3.7954 |
-0.3363 |
0 |
POX33 |
1.293 |
-0.7058 |
0 |
POX55 |
3.796 |
0.355 |
0 |
POX165 |
0.0187 |
-1.3837 |
0 |
|
|
|
|
Dem Phase |
|
|
|
REFL11 |
3 |
179 |
0 |
REFL33 |
165 |
-172 |
0 |
REFL55 |
13 |
170 |
0 |
REFL165 |
86 |
177 |
0 |
AS11 |
-32 |
73 |
0 |
AS33 |
176 |
-72 |
0 |
AS55 |
-41 |
12 |
0 |
AS165 |
-7 |
146 |
0 |
POP11 |
-11 |
-116 |
0 |
POP33 |
124 |
147 |
0 |
POP55 |
-54 |
-146 |
0 |
POP165 |
-117 |
-25 |
0 |
POX11 |
-87 |
15 |
0 |
POX33 |
-105 |
-80 |
0 |
POX55 |
-76 |
16 |
0 |
POX165 |
180 |
-91 |
0 |
PRMI - SB resonant in PRC
SB reso PRMI |
|
|
|
PRCL |
MICH |
SRCL |
REFL11 |
7.6809 |
5.2777 |
0 |
REFL33 |
5.2465 |
3.1565 |
0 |
REFL55 |
7.2937 |
5.589 |
0 |
REFL165 |
4.3892 |
2.6857 |
0 |
AS11 |
1.3123 |
3.545 |
0 |
AS33 |
0.9331 |
1.6022 |
0 |
AS55 |
1.7425 |
4.0514 |
0 |
AS165 |
1.5838 |
1.1344 |
0 |
POP11 |
2.7745 |
0.3791 |
0 |
POP33 |
0.3401 |
-1.7392 |
0 |
POP55 |
2.3872 |
0.6904 |
0 |
POP165 |
-0.5171 |
-2.2279 |
0 |
POX11 |
3.7684 |
1.3574 |
0 |
POX33 |
1.3341 |
-0.7664 |
0 |
POX55 |
3.3815 |
1.6688 |
0 |
POX165 |
0.4785 |
-1.2163 |
0 |
|
|
|
|
Dem Phase
|
|
|
REFL11 |
155 |
-115 |
0 |
REFL33 |
-8 |
3 |
0 |
REFL55 |
91 |
-178 |
0 |
REFL165 |
-62 |
28 |
0 |
AS11 |
109 |
62 |
0 |
AS33 |
-39 |
99 |
0 |
AS55 |
13 |
-38 |
0 |
AS165 |
-155 |
168 |
0 |
POP11 |
141 |
-128 |
0 |
POP33 |
-48 |
-38 |
0 |
POP55 |
24 |
115 |
0 |
POP165 |
95 |
-176 |
0 |
POX11 |
65 |
155 |
0 |
POX33 |
83 |
95 |
0 |
POX55 |
2 |
92 |
0 |
POX165 |
32 |
123 |
0 |
DRMI - SB resonant in PRC
REFL11 |
7.6811 |
5.0417 |
4.2237 |
REFL33 |
5.2751 |
4.1144 |
3.7766 |
REFL55 |
7.2345 |
7.0288 |
6.6801 |
REFL165 |
4.3337 |
4.1266 |
3.7775 |
AS11 |
1.1209 |
3.512 |
0.9248 |
AS33 |
0.9159 |
1.6323 |
0.7971 |
AS55 |
2.6425 |
5.3915 |
2.5519 |
AS165 |
2.6423 |
2.4881 |
2.3272 |
POP11 |
2.7747 |
0.1435 |
-0.6846 |
POP33 |
0.3687 |
-0.7849 |
-1.122 |
POP55 |
2.3244 |
2.1302 |
1.7815 |
POP165 |
-0.5833 |
-0.8 |
-1.1548 |
POX11 |
3.7676 |
3.261 |
0.8086 |
POX33 |
1.3896 |
0.2372 |
0.2333 |
POX55 |
3.4619 |
3.0097 |
3.1326 |
POX165 |
0.782 |
0.6668 |
0.4357 |
|
|
|
|
Dem Phase
|
|
|
REFL11 |
154 |
-16 |
4 |
REFL33 |
-5 |
12 |
51 |
REFL55 |
129 |
-166 |
-123 |
REFL165 |
-23 |
40 |
83 |
AS11 |
132 |
79 |
69 |
AS33 |
-92 |
-127 |
-83 |
AS55 |
-33 |
-55 |
-5 |
AS165 |
154 |
179 |
-144 |
POP11 |
141 |
-29 |
-9 |
POP33 |
-46 |
-27 |
12 |
POP55 |
62 |
127 |
170 |
POP165 |
135 |
-161 |
-117 |
POX11 |
64 |
-102 |
-83 |
POX33 |
85 |
143 |
118 |
POX55 |
57 |
103 |
124 |
POX165 |
99 |
155 |
-164 |
|