Message ID: 5490     Entry time: Tue Sep 20 21:13:39 2011
Author: Suresh 
Type: Update 
Category: IOO 
Subject: MC aligned and PSL beam into MC readjusted 

This morning after Kiwamu maximised the PSL beam coupling into the MC we noticed that the MC2 face camera showed the spot position had moved away from the center by about a diameter.  So I checked the beam spot positions with MCASS and indeed found that the spot on MC2 had moved to about 6mm away from the center in yaw and about 3mm in pitch.  I adjusted the MC2 (and only MC2) to recenter the spots on all the three mirrors.  The new spot positions are given below

spot positions in mm (MC1,2,3 pit MC1,2,3 yaw):
    1.3337   -0.2660    0.6641   -1.0973    0.0468   -1.7130

The PSL beam into MC has been readjusted for maximal coupling into MC.

 

ELOG V3.1.3-