Entry  Mon Sep 19 23:19:40 2011, Keiko, Update, IOO, AM modulation mistery 
    Reply  Tue Sep 20 23:01:37 2011, Keiko, Update, IOO, AM modulation mistery 
       Reply  Wed Sep 21 16:44:18 2011, Keiko, Update, IOO, AM modulation mistery P9210138.JPGP92101381.jpg
          Reply  Wed Sep 21 17:42:35 2011, rana, Update, IOO, AM modulation misery 
             Reply  Wed Sep 21 18:53:03 2011, Keiko, Update, IOO, AM modulation misery AM0.pngAMratio1.pngAMratio50.png
                Reply  Wed Sep 21 21:13:35 2011, rana, Update, IOO, AM modulation misery 
                   Reply  Thu Sep 22 17:29:42 2011, Keiko, Update, IOO, AM modulation mistery AMmod0MICH.pngAMmod1e-1MICH.pngAMmod0SRCL.pngAMmod3e-3SRCL.pngAMmod3e-3SRCL-AS55I.png
                      Reply  Sat Sep 24 09:55:42 2011, Keiko, Update, IOO, AM modulation mistery 
Message ID: 5472     Entry time: Mon Sep 19 23:19:40 2011     Reply to this: 5491
Author: Keiko 
Type: Update 
Category: IOO 
Subject: AM modulation mistery 

 Keiko, Anamaria

We started to investigate the AM modulation mistery again. Checking just after the EOM, there are AM modulation about -45dBm. Even if we adjust the HWP just before the EOM, AM components grow up in 5 mins. This is the same situation as before. Only the difference from before is that we don't have PBS and HWP between the EOM and the monitor PD. So we have a simpler setup this time.

We will try to align the pockells cell alignment tomorrow daytime, as it may be a problem when the crystal and the beam are not well parallel. This adjustment has been done before and it didn't improve AM level at that time.